Catalog Number 0250080233
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/02/2015
Event Type  malfunction  
Manufacturer Narrative
The button screw cover cap located on the proximal end of the device is missing.Also it was noted cracks, dents and scratches where seen on insulation.Functional inspection : the instrument failed the insul scan.The probable root causes are normal wear, user misuse and improper sterilization methods.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
 
Event Description
It was reported that a cap on the screw below the push putton came off.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5MM MONOPOLAR HANDLE 33CM
Type of Device
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
kimberly lynch
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key5109505
MDR Text Key27063708
Report Number0002936485-2015-00859
Device Sequence Number1
Product Code OCZ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K973259
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Report Date 07/31/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250080233
Device Lot Number1340503D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2015
Date Manufacturer Received09/02/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-