BIOMET ORTHOPEDICS COMP LK SCR 3.5HEX 4.75X35 ST; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problems Fracture (1260); Fitting Problem (2183)
Patient Problem Unspecified Infection (1930)
Event Date 09/03/2015
Event Type  Injury  
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.There are warnings in the package insert that state that this type of event can occur: under warnings, it states, ¿improper selection, placement, positioning, alignment and fixation of the implant components may result in unusual stress conditions which may lead to subsequent reduction in the service life of the prosthetic components.¿ this report is number 1 of 2 mdrs filed for the same event (reference 1825034-2015-04119 / 04120).
 
Event Description
It was reported that the patient had an initial shoulder procedure on an unknown date.Subsequently, the patient was revised due to infection on an unknown date.The patient underwent another shoulder procedure on (b)(6) 2015.During the procedure, the poly would not fit into the baseplate as the locking screws were sitting too proud in the baseplate.The surgeon removed the screws and repositioned a few times.The procedure was completed with shorter screws.This caused a delay of 30 minutes.Additionally, while impacting the poly the surgeon fractured the impactor and pieces fell into the patient.The fracture pieces we retrieved from the patient.
 
Manufacturer Narrative
This follow-up report is being filed to relay additional information, which was unknown at the time of the initial medwatch.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP LK SCR 3.5HEX 4.75X35 ST
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
megan haas
56 e. bell drive
warsaw, IN 46582
5743726700
MDR Report Key5112211
MDR Text Key27057428
Report Number0001825034-2015-04119
Device Sequence Number1
Product Code PAO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK132239
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Followup
Report Date 09/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number180554
Device Lot Number329580
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/30/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/03/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age54 YR
Patient Weight132
-
-