• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASCENSIA DIABETES CARE US INC. CONTOUR NEXT; QUALITY CONTROL MATERIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSIA DIABETES CARE US INC. CONTOUR NEXT; QUALITY CONTROL MATERIAL Back to Search Results
Model Number 7314
Device Problems Device Expiration Issue (1216); Incorrect, Inadequate or Imprecise Result or Readings (1535); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/01/2015
Event Type  malfunction  
Event Description
The customer ran a test on the contour next ez with expired control solution.The meter did not automatically mark it as a control test, which will be displayed as a blood result when accessing the meter's memory.No adverse event was alleged.Customer was advised to return the control for evaluation.New meter, strips and control were sent to the customer.
 
Manufacturer Narrative
See remedial action and correction/removal reporting number.This information was not provided in the initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR NEXT
Type of Device
QUALITY CONTROL MATERIAL
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
430 s. beiger st.
mishawaka IN 46544
Manufacturer (Section G)
BAYER HEALTHCARE LLC
430 s. beiger st.
mishawaka IN 46544
Manufacturer Contact
jeri messmore
430 s. beiger st.
mishawaka, IN 46544
5742567719
MDR Report Key5113198
MDR Text Key27212858
Report Number1826988-2015-00498
Device Sequence Number1
Product Code JJX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111268
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Remedial Action Recall
Type of Report Followup
Report Date 09/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/30/2014
Device Model Number7314
Device Lot Number2697
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received09/30/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction NumberZ-2482-2015