(b)(4).Method: two complaint rt266 infant dual heated evaqua2 breathing circuits were returned to fisher & paykel healthcare in (b)(4) for evaluation.The circuits were visually inspected and pressure tested.Results: visual inspection revealed that the collar at the patient end of one of the two circuits, at the expiratory limb, was cracked.Pin holes were observed in the inspiratory limb, approximately 620 mm from the chamber end connector.The pressure test revealed that this circuit was out of specification.There was no visible damage found to the other circuit.Pressure test revealed that the returned breathing circuit was within specification and did not exhibit excessive leak.A lot check was not performed as lot information was not provided.Conclusion: based on the nature of the observed cracking and previous investigations into this type of failure, the cracking is most likely due to the connector coming into contact with a cleaning chemical, resulting in environmental stress cracking.However, we were unable to determine definitively the root cause of the observed pin holes in the inspiratory limb.All rt266 infant dual heated evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.The healthcare facility reported that the damage was noticed while the patient was setup with the circuit, which suggests that the subject breathing circuit became damaged after the product was released for distribution.Our user instructions that accompany the rt266 infant dual heated evaqua2 breathing circuit state the following: - "check all connections are tight before use." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "set appropriate ventilator alarms." - "do not soak, wash, sterilize or reuse this product.Avoid contact with chemicals, cleaning agents or hand sanitizers.".
|