(b)(4).The device history record for this lot number was reviewed.100% visual inspection for broken or missing parts and pin alignment is performed on each assembly during the manufacturing process.The ring separation force and the ring retention force is within specification.The doppler probes are inspected 100% for visual defects and tested 100% for functionality.According to the device history record, the device met specification prior to market release.The flow-coupler doppler probe wire was cut by the surgeon when the patient was taken back into surgery.The doppler probe was discarded and a portion of the probe wire was returned to synovis for evaluation.The external lead wire was also returned for evaluation.There were no marks that would indicate damage to probe wire.There was no functional testing performed with the returned portion of the probe wire.The external lead wire was tested and found to be functioning as intended.With the probe wire cut by the surgeon and the doppler probe discarded, a full investigation cannot be completed and root cause of the lost signal cannot be determined.
|