Actual device returned to richard wolf medical instrumentation corporation (rwmic) for investigation on (b)(6) 2015.Device missing approximately 15mm of insulation from distal end of electrode, appears to have been cut.Device was sent in with the sheath that was being used during the time of the issue.The sheaths ceramic tip had pieces broken off from the distal tip leaving sharp ends which could have cut the insulation from this device.This device was tested with hypot tester and no electrical short found.Requests for missing/additional information have been submitted, no response as of (b)(6) 2015.Manufacture date: 04/07/2015.Purchase date: (b)(6) 2015.No similar issues on this device on the last three years.Labeling was reviewed and found to be adequate.Ie intended use, indications and field of use, preparation and cautions regarding high temperatures and use of irrigation fluid.Rwmic considers this matter closed.However, in the event rwmic receives additional information, a follow-up report will be provided to fda.
|
Facility reported to one of richard wolf medical instruments corporation (rwmic) customer repair specialist that a device arched and damaged the tip of a second device being used during the procedure.The two impacted devices were swapped out to back up devices, readily available for use, and procedure was completed as scheduled.No injury to patient or staff reported.At least six devices are assembled together in order to operate.The two devices which had product problems and their associated mdr's are: sheath (8661.3741) - mdr 1418479-2015-00031 (tip damaged).Electrode (8424.141) - mdr 1418479-2015-00032 (device arched).
|