Catalog Number 5100004000 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/11/2015 |
Event Type
malfunction
|
Event Description
|
The device was returned to stryker instruments for service.During functional testing by a service technician, it was found that the device had a run on.No medical intervention and no adverse consequences were reported with this event.As this event occurred during testing at the manufacturer facility, there was no patient involvement and no delay to a surgical procedure.
|
|
Manufacturer Narrative
|
The technician duplicated run-on with the device.A broken or worn cable is known to contribute or cause the failure.The device was scrapped by the manufacturer.
|
|
Event Description
|
The device was returned to stryker instruments for service.During functional testing by a service technician, it was found that the device had a run on.No medical intervention and no adverse consequences were reported with this event.As this event occurred during testing at the manufacturer facility, there was no patient involvement and no delay to a surgical procedure.
|
|
Search Alerts/Recalls
|