(b)(4).Date of follow-up report : 11/03/2015.One used lf1212 was received for evaluation.Visual inspection found the jaw melted at the tip.The customer reported the insulating material on the tip was found partially broken.The investigation found the tip of one of the electrode jaws was charred and melted.No piece appeared to be broken off but had melted and exposed a portion of the metal on the tip of the jaw.The tip may have contacted a metal instrument causing an increase in current flow.This is an isolated incident, no other complaints of this nature have been reported.The investigation identified the root cause of the reported event to be attributed to user technique or contacting a metal instrument.The instructions for use cautions the user to not use the lf1212 as a bipolar scissors.Continual activation to separate tissue promotes eschar buildup on seal surfaces that can reduce the effectiveness of the seal and may create a condition that compromises the jaws insulating coating.The ifu warns, contact between an active instrument electrode and any metal object (hemostats, staples, clips, retractors, etc.) may increase current flow and may result in unintended surgical effects.No trend has been identified for this failure mode.
|