• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TAEWOONG MEDICAL CO., LTD NITI-S ESOPHAGEAL COMVI STENT ESOPHAGEAL STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TAEWOONG MEDICAL CO., LTD NITI-S ESOPHAGEAL COMVI STENT ESOPHAGEAL STENT Back to Search Results
Model Number EC1815AR
Device Problems Unraveled Material (1664); Activation Failure (3270)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/28/2015
Event Type  malfunction  
Manufacturer Narrative
Since the device was not returned, it is impossible to proceed to actual evaluation.But, it is confirmed from the device history record that the suspected device had been manufactured with no significant issue and passed all the inspections.It is hard to find out exact root cause for this complaint, because the suspected device was not returned and it is difficult to reconstruct the situation at that time in the lab and ltd info.We will continuously monitor whether similar or same complaint occurs.The suspected device is not registered to us fda and it has not been shipped into the us.
 
Event Description
The ec1815ar was applied to cardiac stenosis.After stent deployment, stent moved up to thoracic esophagus as the tip of delivery system caught the stent cell.Because proximal end stent cell was broken, physician removed it once and ec1812ar was newly implanted to finish the procedure.Reportedly, physician did not push back the outer sheath to its original position before removal.There were no patient complications as a result of this event.
 
Manufacturer Narrative
It was reported that stent cell was broken based on complaint description received.However, no cell was broken and it was only loosen as a result of our analysis with returned device.It is impossible to identify the exact root cause since it is hard to recreate the situation at the time of procedure.However, it is considered that stent cell was loosen since removal of delivery system was tried after deployment without placing inner sheath back into outer sheath by user based on complaint description.It is specified on our user manual as follows in order to prevent these problems.It should be careful when removing the introducer system and guidewire after stent deployment.And if it is felt excessive resistance during removal of them, wait 3-5 minutes to allow further stent expansion.(place the inner sheath back into the outer sheath as the original state prior to removal.)" also no defect on device history record was found since 100% sampling inspection was performed during process inspection in order to check tip was properly adhered on its inner sheath.It will be monitored that similar case is occurred from now on.The suspected device is not registered to us fda and it has not been shipped into the us.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NITI-S ESOPHAGEAL COMVI STENT ESOPHAGEAL STENT
Type of Device
ESOPHAGEAL STENT
Manufacturer (Section D)
TAEWOONG MEDICAL CO., LTD
gimpo-si, gyeonggi-do
KS 
Manufacturer Contact
14, gojeong-ro, wolgot-myeon
gimpo-si, gyeonggi-do 
19960641
MDR Report Key5137855
MDR Text Key27920848
Report Number3003902943-2015-00068
Device Sequence Number1
Product Code ESW
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Not Applicable
Type of Report Followup
Report Date 09/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/19/2017
Device Model NumberEC1815AR
Device Catalogue NumberTW-TC-01
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer10/02/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received09/08/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-