It was reported that during a cryoablation procedure, after insertion of the mapping catheter into the right inferior pulmonary vein (ripv), the catheter could not be retracted out of the pulmonary vein (pv).It was noted that the catheter was torn; the physician then decided to cut the end and retract it over the wire with a catheter.The mapping catheter was retracted successfully and the procedure was aborted for safety reasons.Further follow-up indicated that a dissection of the pulmonary vein was noticed when the event occurred, and that the catheter appeared kinked after the successful retraction.No further patient complications have been reported as a result of this event.
|
Product event summary: data files and the device, mapping catheter 990063-015 with lot 209411791, were returned and analyzed.The data files did not show any system notices on the reported date of the event.Upon return, the catheter was inside a competitor sheath for capturing the loop array.Visual inspection of the catheter showed the catheter was broken at the proximal end attachment with the electrocardiogram (ecg) connector; all ecg signal wires were broken inside the shaft.It was noted that this was broken post procedure to introduce the mating device to capture the catheter¿s loop section.Additionally, the loop array was misshapen and the array was damaged; the covering in the array section appeared twisted counterclockwise, in scope with the instructions for use.The pebax tubing was ribbed at multiple locations along the pebax starting at the transition with the metal shaft to the loop array; all eight electrodes exist on the loop section.These product analysis findings do not indicate a manufacturing related defect; the damaged tip/loop section and kink were caused by excessive maneuvering of the catheter to retract it from the entrapment of the pulmonary vein.The dissection of the pulmonary vein and entrapment of the catheter in the heart are clinical issues encountered during the procedure.In conclusion, the kink and damaged loop array of the mapping catheter were confirmed through visual inspection.The mapping catheter failed the return product inspection due to the damaged tip/loop section, kink in the tip/loop section, ribbed pebax tubing, and broken shaft at the ecg connector attachment.Clinical issues of catheter entrapment in the heart and pulmonary vein dissection were encountered during the procedure.
|