• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1125275-15
Device Problems Kinked (1339); Failure to Advance (2524); Material Deformation (2976); Torn Material (3024)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/11/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.(b)(4).Evaluation summary:the device was returned for analysis.The guide wire exit notch was torn and the outer member was torn at the guide wire exit notch.The reported stent damage and kinks were able to be confirmed.The tearing of the guide wire exit notch appears to have resulted from an interaction with the guide wire.The failure to advance could not be replicated in a testing environment as it was based on operational circumstances.Based on a visual and dimensional inspection of the returned device, there is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event Description
It was reported that during a procedure of the eccentric, heavily tortuous, heavily calcified, 99% stenosed, proximal circumflex artery after pre-dilatation with a 2.0 x 15 mm tenku balloon dilatation catheter (bdc) the 2.75 x 15 mm xience alpine stent delivery system (sds) was advanced to the lesion but failed to cross.The device was removed from the anatomy without reported issue and it was noted that the shaft was kinked at several locations and the stent struts were found flared.A non-abbott stent was used in the procedure.There was no reported adverse patient effect.There was no reported clinically significant delay in the procedure.No additional information was provided.Return device analysis revealed the guide wire exit notch was torn and the outer member was torn at the guide wire exit notch.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key5138684
MDR Text Key27977500
Report Number2024168-2015-05925
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Health Professional
Remedial Action Other
Report Date 10/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/08/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/01/2018
Device Catalogue Number1125275-15
Device Lot Number5012141
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/14/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-