Siemens current product support (cps) team investigated customer returned data.For one particular patient which printouts of the rp500 and the abl835 results were provided, the customer noted that the nbili result was '<2.0' on the rp500 but '16.6' on the abl835.In actuality, this sample was flagged on the rp500 as questionable results (with ? marks) for thb, the co-ox fractions, and nbili.Based on the flagging message, the customer should not have acted on these results.There is indication that some of the specimens may have varying degrees of hemolysis, based on elevated k+ results.A robust side-by-side method comparison of nbili between the rp500 and the abl835 (or the lab chemistry system) would have to be conducted at the hospital site on at least 40 paired patient specimens in order to understand the true relationship between the two systems.Similar studies of comparable siemens blood gas systems have not shown the trend indicated by this customer (on very limited dataset).If a method correlation is established, the use of an appropriate set of correlation coefficients values can be utilized so that the rp500 nbili assay results more closely matches the comparative device (be it the abl or any chemistry analyzer model, for instance).This correlation coefficient feature is offered in system set-up.Evidence available indicates that the rp500 co-oximeter (which dictates thb, fractions and nbili performance) was functioning correctly.There is no evidence that the rp500 nbili is not performance to expectations.
|