CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problem Operating System Becomes Nonfunctional (2996)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/14/2015
Event Type  malfunction  
Manufacturer Narrative
 
Event Description
A company representative reported that a handheld device was not working.He performed troubleshooting and stated that the device would freeze upon interrogation.This happened with multiple patients, but all patient devices were eventually interrogated.The handheld device, software flashcard, and programming wand have been returned to the manufacturer where analysis is currently underway.
 
Manufacturer Narrative
Corrected data: new information corrects the suspect device.
 
Event Description
Analysis of the handheld was completed on 11/03/2015.During the analysis, it was identified that the handheld would not power on.The cause for the anomaly is associated with a swollen main battery.The swollen battery bent the battery cover causing the battery latch switch to register that the latch was unlocked, thus preventing the handheld from powering on (this condition can also cause the handheld to power off intermittently, but cannot cause the handheld to freeze while in use).No further anomalies were identified.Analysis of the software was completed on 11/03/2015.No anomalies associated with flashcard software were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5146319
MDR Text Key28514839
Report Number1644487-2015-06104
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1075378
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2015
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received10/12/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/15/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-