The cause for the discordant advia centaur xp psa results is unknown.Siemens healthcare diagnostics has requested the patient samples for further testing and investigation.The instrument is performing within specification.The ifu states in the limitations section: "note: do not interpret levels of psa as absolute evidence of the presence or the absence of malignant disease.Before treatment, patients with confirmed prostate carcinoma frequently have levels of psa within the range observed in healthy individuals.Elevated levels of psa can be observed in patients with nonmalignant diseases.Measurements of psa should always be used in conjunction with other diagnostic procedures, including information from the patient's clinical evaluation." (b)(4).
|
Siemens filed the initial mdr 1219913-2015-00156 on october 14, 2015.12/20/2015 additional information: siemens received two samples for testing and investigation.The patient samples were (b)(6).The patient samples were tested with lot 263 for the high dose hook effect claim.Summary of results: sid: (b)(6).Units: ng/ml , ng/ml.Total psa, neat: > 100 (106) , 74.8.Total psa, diluted (1:100): 6075.5, 11427.0.Free psa : 435.3, 1855.8.Cpsa: 6557.2, 8839.9.% free psa (fpsa/total psa): 7%, 16%.% cpsa (cpsa/total psa): 108%, 77%.Ifu hook claim - pass/fail: pass, fail.The patient sample (b)(6) did not hook back onto the curve.The value was > 100 ng/ml and did not confirm for the high dose hook effect.The patient sample (b)(6) did hook back onto the curve and confirmed for the high dose hook effect as it recovered <100 when tested neat.The value was 74.8 ng/ml.When the patient sample was diluted, the concentration was calculated to be 11,427 ng/ml.The ifu states in the high dose hook effect section: "patient samples with high total psa levels can cause a paradoxical decrease in the rlus (high-dose hook effect).In this assay, patient samples with total psa levels as high as 10,000 ng/ml (10,000 ug/l) will assay greater than 100 ng/ml (100 ug/l)." siemens has also performed an internal investigation for the advia centaur systems psa high dose hook effect.Based on siemens' internal testing, samples with total psa concentrations between 4200ng/ml (ug/l) and 8400 ng/ml (ug/l) do not assay >100 ng/ml (ug/l) but instead result in falsely depressed concentrations of approximately 50 ng/ml (ug/l) to 94 ng/ml (ug/l).Therefore, siemens issued urgent field corrective action cc 16-05.A.Ous and urgent medical device correction cc 16-05.A.Us on december 2015 informing the customer that the advia centaur systems prostate specific antigen (psa) assay kit lots ending in 263 and above are not meeting the current high-dose hook effect expectation in the instructions for use (ifu).Instructions on actions to be taken by the customer are provided in the customer communication.No further action required.
|