Model Number 16-02-80
Device Problem Contamination (1120)
Patient Problem No Patient Involvement (2645)
Event Date 09/15/2015
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater/cooler 16-02-80 is not released in the us, but it is a variant of the heater/cooler 16-02-85, which is released in the us ((b)(4)).Sorin group (b)(4) manufactures the sorin heater/cooler system 3t.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the sorin heater/cooler system 3t was tested positive for mycobacteria.There was no patient involvement.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that the sorin heater/cooler system 3t was tested positive for mycobacteria.There was no patient involvement.
 
Manufacturer Narrative
Sorin implemented a field safety notice for disinfection and cleaning of sorin heater cooler devices.The z number is z-2076/2081-2015.Sorin group (b)(4) manufactures the sorin heater-cooler system 3t.The incident occurred in (b)(6).(b)(4).Sorin group (b)(4) received a report that the sorin heater-cooler system 3t system tested positive for mycobacteria.There was no report of any patient involvement.A sorin group service technician was dispatched to the facility to investigate the device.A visual inspection of the outer and inner parts of the heater-cooler unit ((b)(4)) revealed minor residuals and biofilm in several locations including the pumps and water circuits.Based on these findings, it was concluded that this issue was the result of the user failing to adhere to the cleaning and disinfection instructions outlined in the ifu.The customer has expressed no intention of sending the unit to sorin group (b)(4) for disinfection.The unit has been returned to service.As corrective action, fsca 9611109-06/03/15-002-c was released to remind our customers about the importance of adhering to the water management and disinfection procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS, PRODUCT CODE: DWC
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND GMBH
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND GMBH
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
carrie wood
14401 w. 65th way
arvada, CO 80004
3034676461
MDR Report Key5150404
MDR Text Key28650832
Report Number9611109-2015-00460
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Device Catalogue NumberN/A
Device Lot NumberN/A
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/22/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/12/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-