• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT1 ANALYZER, IMMUNO READY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT1 ANALYZER, IMMUNO READY Back to Search Results
Catalog Number 04P75-01R
Device Problems Circuit Failure (1089); Device Emits Odor (1425); Smoking (1585); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/08/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
On 10/08/2015, abbott point of care (apoc) was contacted by a customer who reported when a new pack was inserted and the analyzer turned on, crackling and popping sounds were heard.The analyzer began to smoke and there is a lingering burnt smell.Apoc has determined that a component failure within the analyzer circuitry, may lead to the batteries becoming uncomfortably hot to touch in the area of the battery compartment when using a green non-fused battery carrier.However, the customer states that rechargeable batteries and also a red fused battery carrier was being used.Therefore the analyzer is unlikely to become hot to touch.The product was replaced and returned for investigation.Based on the information available, there were no patient or user related injuries associated with this complaint.
 
Manufacturer Narrative
(b)(4).The investigation was completed on (b)(6) 2015.The failure was due to a defective tantalum capacitor.
 
Event Description
Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT1 ANALYZER, IMMUNO READY
Type of Device
I-STAT1 ANALYZER
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key5155505
MDR Text Key28908722
Report Number2245578-2015-00058
Device Sequence Number1
Product Code CGA
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K001387
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/16/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04P75-01R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/06/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/23/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-