SYNTHES OBERDORF OSCILLATING SAW ATTACHMENT II; BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL
|
Back to Search Results |
|
Catalog Number 511.801 |
Device Problem
Partial Blockage (1065)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event was unknown.(b)(6).The actual device was returned for evaluation.During engineering evaluation of the device it was determined that the gear blocked, seized and rough running.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If information is obtained that was not available for the initial medwatch, a supplemental medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported from (b)(6) that the locking part on the oscillating saw attachment device was defective.During in-house engineering evaluation, it was observed that the gear blocked, seized and rough running.It was not reported if the device was used in surgery, or if there was patient involvement.There were no delays to a scheduled surgical procedure.A spare device was not available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|