MAUDE Adverse Event Report: ATRIUM MEDICAL CORPORATION ATRIUM ICAST COVERED STENT; PTFE COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ATRIUM MEDICAL CORPORATION ATRIUM ICAST COVERED STENT; PTFE COVERED STENT Back to Search Results
Model Number 85440
Device Problem Device Slipped (1584)
Patient Problem No Information (3190)
Event Date 09/30/2015
Event Type  Injury  
Manufacturer Narrative
On completion of the investigation a follow up report will be submitted.
 
Event Description
The stent slipped off the balloon as it was advanced into the sheath.
 
Manufacturer Narrative
Engineering analysis: the complaint details provided indicate that the stent dislodged off the balloon within the introducer sheath.The returned device balloon surface was evaluated to determine if the stent was crimped properly during manufacturing.The crimped stent impressions were clearly visible indicating that the stent was properly crimped during manufacturing.The crimped stent diameter was measured and was 2.14mm.This diameter when compared to the lot qualification test data and falls within the same range of diameters that are only slightly smaller.The slightly larger diameter is due to the stent being pulled over the folded balloon cones.The introducer sheath used in the case was not returned.Engineering summary: a full review of the catheter lot history records for the device in question was performed.The records indicate that this lot of catheters passed atriums final lot qualification testing.This inspection requires that the catheter lot must pass the following: · ability of the stent and delivery system to be passed through the labeled introducer sheath.· ability to deploy the stent at nominal pressure (8atm).· ability to withdraw the deflated balloon catheter back through the labeled introducer sheath · ability of the delivery system to withstand 10 inflate/deflate cycles at the rated burst pressure (12atm) without leaks or failures.· balloon burst testing.The balloon must burst over the rated burst pressure specified on the label (12atm) · manifold to shaft tensile testing.Samples when tensile tested must not break or separate below 3.37lbs result: all 59 quality inspection samples passed this final inspection without any non-conformances noted during the final lot qualification testing related to the stent.All samples passed through the introducer sheath without any stent dislodgments.During the final lot qualification data shows that all 59 test samples were able to pass through the 6fr introducer sheath without issue.Since december of 2013 over 30,000 test units have been passed through the introducer sheath without a failure for the inability of the stent to pass through the introducer sheath.The product in question has also been subjected to simulated use in a tortuous iliac artery model whereas the stent delivery system is advanced contra laterally over the iliac arch through a 6fr 55cm long cook check flow performer introducer sheath.The stent is then deployed at nominal pressure as specified on the product label and the balloon deflated and withdrawn back through the introducer sheath.This testing was conducted numerous times while being submerged in a heated water bath at 37°c (body temperature) during design verification testing of the product.None of the samples tested had an issue regarding stent dislodgments while passing through the introducer sheath.Conclusion: based on the details of the event and the successful lot qualification test data, atrium can find no fault with the device and or lot of stent delivery systems in question.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATRIUM ICAST COVERED STENT
Type of Device
PTFE COVERED STENT
Manufacturer (Section D)
ATRIUM MEDICAL CORPORATION
5 wentworth drive
hudson NH 03051
Manufacturer (Section G)
ATRIUM MEDICAL CORPORATION
5 wentworth drive
hudson NH 03051
Manufacturer Contact
lynda mclaughlin
40 continental blvd.
merrimack, NH 03054
6038645470
MDR Report Key5155727
MDR Text Key28457283
Report Number1219977-2015-00305
Device Sequence Number1
Product Code JCT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050814
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/16/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Expiration Date07/31/2017
Device Model Number85440
Device Catalogue Number85440
Device Lot Number211656
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/02/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/03/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age57 YR
-
-