CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problem Operating System Becomes Nonfunctional (2996)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/21/2015
Event Type  malfunction  
Event Description
It was reported that the physician's programming system was freezing when trying to interrogate patients recently.The physician tried to perform hard resets which did not resolve the issue.The physician indicated that this did not affect programming the patient's device as he was only trying to interrogate.So the patient's therapy was not affected.The physician was provided a new programming tablet.The handheld is expected to be returned for analysis, but has not been received to date.
 
Event Description
The handheld and flashcard were received for analysis.Analysis is underway, but has not been completed to date.
 
Event Description
Analysis of the handheld was completed on 11/17/2015.The cause for the display anomaly is associated with resistance values being higher than expected in the touch screen circuitry.Once the display was replaced with a known good display, no further anomalies associated with the handheld performance were identified during the analysis.Analysis of the flashcard was completed on 11/17/2015.No anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5156461
MDR Text Key28922767
Report Number1644487-2015-06149
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/16/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMODEL 250
Device Lot Number1075841
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2015
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received11/17/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/16/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-