• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS 8000 CARDIOPLEGIA MONITOR; ALARM, BLOOD-PRESSURE-SARNS 8000 CARDIOPLEGIA MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS 8000 CARDIOPLEGIA MONITOR; ALARM, BLOOD-PRESSURE-SARNS 8000 CARDIOPLEGIA MONITOR Back to Search Results
Model Number 16414
Device Problems No Display/Image (1183); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/01/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The subsidiary site service engineer (se) found ampro board was broken and caused the reported problem.The se assembled an old model central processing unit (cpu) board, and now the cpg monitor is working correctly.Per the se, they are not going to send the defective printed circuit board assembly (pcba) to the manufacturer for evaluation.He stated there was no damage or obvious component damage to board.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the cardioplegia (cpg) monitor is broken.At first, the monitor housing was very hot after running for some time.Sometimes the numbers on the display of the monitor were frozen.When the user presses the buttons, there's no affect.Now the monitor does not boot-up and always has black screen.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed via subsidiary evaluation.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS 8000 CARDIOPLEGIA MONITOR
Type of Device
ALARM, BLOOD-PRESSURE-SARNS 8000 CARDIOPLEGIA MONITOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5156684
MDR Text Key28910180
Report Number1828100-2015-00896
Device Sequence Number1
Product Code DSJ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K960916
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/16/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16414
Device Catalogue Number16414
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received12/22/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/13/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-