• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY IMPLANTS (A DIVISION OF DENTSPLY IH AB) IMPLANT DRIVER; ACCESSORIES, IMPLANT, DENTAL, ENDOSSEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY IMPLANTS (A DIVISION OF DENTSPLY IH AB) IMPLANT DRIVER; ACCESSORIES, IMPLANT, DENTAL, ENDOSSEOUS Back to Search Results
Device Problems Positioning Failure (1158); Fracture (1260); Material Torqued (2980)
Patient Problem Failure of Implant (1924)
Event Date 08/03/2015
Event Type  Injury  
Manufacturer Narrative
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.A shaft fragment from a driver stuck in the handle, which is the reason, that no other handle can be fixed.The fragment is not at all in the position a shaft usually should be, when placed in the handle correct and as far up as described in the ifu.It was torn about 90° and hereby bent back the two clamping lashes inside the driver.
 
Event Description
In this event it was reported that an ev driver was stripped and would not engage.The session was not completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANT DRIVER
Type of Device
ACCESSORIES, IMPLANT, DENTAL, ENDOSSEOUS
Manufacturer (Section D)
DENTSPLY IMPLANTS (A DIVISION OF DENTSPLY IH AB)
aminogatan 1
molndal, vastra gotalands lan [se-14] S-431 21
SW  S-431 21
Manufacturer (Section G)
DENTSPLY IMPLANTS (A DIVISION OF DENTSPLY IH AB)
aminogatan 1
molndal, vastra gotalands lan [se-14] S-431 21
SW   S-431 21
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key5157639
MDR Text Key28556097
Report Number9612468-2015-00022
Device Sequence Number1
Product Code NDP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 08/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/16/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Dentist
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/10/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-