• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS PLATINUM 1 SERIES; SURGICAL ADJUNCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS PLATINUM 1 SERIES; SURGICAL ADJUNCTS Back to Search Results
Model Number 1MTEC30
Device Problems Crack (1135); Positioning Failure (1158)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/27/2015
Event Type  malfunction  
Manufacturer Narrative
As lens was partially delivered into the eye and not fully implanted.All pertinent information available to abbott medical optics has been submitted.
 
Event Description
It was reported that when attempting to insert the intraocular lens (iol) the lens started coming through a weak area in the plastic of the cartridge.The report indicates the lens touched the eye but never fully was in the eye and got stuck in the delivery device.There was no patient injury reported and the issue was not attributed to an account use error.No other information was provided.
 
Manufacturer Narrative
The complaint sample (1mtec30 cartridge) was not received.Only the lens was received.The reported device problem code of cartridge crack could not be verified.Manufacturing record review: the manufacturing record review could not be performed because the lot number for the cartridge is unknown.Labeling review: the directions for use (dfu) were reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATINUM 1 SERIES
Type of Device
SURGICAL ADJUNCTS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
Manufacturer (Section G)
ABBOTT MEDICAL OPTICS INC.
road 402 north, km 4.2
anasco industrial park, pob 14
anasco PR 00610
Manufacturer Contact
pamela mcclain
1700 east st. andrew place
santa ana, CA 92705
7142478243
MDR Report Key5159531
MDR Text Key28609325
Report Number2648035-2015-00995
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081545
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model Number1MTEC30
Device Catalogue Number1MTEC30
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/31/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-