• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/25/2015
Event Type  malfunction  
Manufacturer Narrative
 
Event Description
It was reported that physician's handheld screen was too dim to see.The physician was provided a new programming tablet.The handheld is expected to be received for analysis, but has not been received to date.
 
Event Description
The handheld and flashcard were received for analysis.Analysis of the handheld was completed on 12/17/2015.During the analysis, no anomalies associated with the screen brightness was identified.During the analysis, it was identified that the handheld would not power on.The cause for the anomaly is associated with a swollen main battery.The swollen battery bent the battery cover causing the battery latch switch to register that the latch was unlocked, thus preventing the handheld from powering on (this condition can also cause the handheld to power off intermittently).No further anomalies were identified.Analysis of the flashcard was completed on 12/17/2015.No anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5162649
MDR Text Key28986977
Report Number1644487-2015-06180
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/25/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1073772
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2015
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received12/01/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/27/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-