The handheld and flashcard were received for analysis.Analysis of the handheld was completed on 12/17/2015.During the analysis, no anomalies associated with the screen brightness was identified.During the analysis, it was identified that the handheld would not power on.The cause for the anomaly is associated with a swollen main battery.The swollen battery bent the battery cover causing the battery latch switch to register that the latch was unlocked, thus preventing the handheld from powering on (this condition can also cause the handheld to power off intermittently).No further anomalies were identified.Analysis of the flashcard was completed on 12/17/2015.No anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
|