Engineering analysis: the complaint details provided indicate that the catheter shaft of the stent delivery system buckled due to the tortuosity of the anatomy as specified above in the complaint details.The returned device was evaluated to determine the cause of the complaint.The device was removed from the package and the stent was still properly secured to the folded balloon.The catheter shaft was kinked approximately 10cm from the manifold strain relief and also mid shaft and again at the proximal balloon bond.The shaft appears to have been compromised by the use of excessive force due to the tortuosity of the patient's anatomy.The instructions for use (ifu) specify the following: "caution: if resistance is encountered, do not force passage." the stent was still situated between the two radiopaque marker bands and was securely crimped on to the balloon.The crimped stent diameter was measured and was 2.41mm.This diameter when compared to the lot qualification test data falls within the same range of diameters.The introducer sheath used in the case was not returned.The stent delivery system was prepped per the instructions for use and brought to 8atm as specified on the product label.The stent deployed properly and without issue.Engineering summary: a full review of the catheter lot history records for the device in question was performed.The records indicate that this lot of catheters passed atriums final lot qualification testing.This inspection requires that the catheter lot must pass the following: ability of the stent and delivery system to be passed through the labeled introducer sheath.Ability to deploy the stent at nominal pressure (8atm).Ability to withdraw the deflated balloon catheter back through the labeled introducer sheath ability of the delivery system to withstand 10 inflate/deflate cycles at the rated burst pressure (12atm) without leaks or failures.Balloon burst testing.The balloon must burst over the rated burst pressure specified on the label (12atm).Manifold to shaft tensile testing.Samples when tensile tested must not break or separate below 3.37lbs.Result: all 59 quality inspection samples passed this final inspection without any non-conformances noted during the final lot qualification testing related to the stent.Conclusion: based on the details of the event and the successful lot qualification test data, atrium can find no fault with the device and or lot of stent delivery systems in question.
|