Model Number VTICMO12.1
Device Problem Crack (1135)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/20/2015
Event Type  malfunction  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.Method - (process evaluation): work order search.Results - (evaluation result): a lens work order search was performed and no similar complaints were found within the work order.Conclusion - (unable to confirm complaint): based on the complaint history and work order search, a specific root cause of the event could not be determined.(b)(4).
 
Event Description
The reporter indicated the surgeon inserted a 12.1mm vticmo12.1 implantable collamer lens, -8.5/+2.0/096 diopter, in the patient's right eye (od) and the lens was noted to have a crack upon insertion.There was no reported patient injury and the lens was exchanged for another same model lens and the problem was resolved.The patient's post-op best-corrected visual acuity was 20/20.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICL (IMPLANTABLE COLLAMER LENS)
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
althea watson
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key5166980
MDR Text Key28910194
Report Number2023826-2015-01376
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial
Report Date 09/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date03/31/2018
Device Model NumberVTICMO12.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/23/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/09/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age22 YR
-
-