The customer reported that the companion 2 driver exhibited a left high pressure alarm while supporting a patient.The customer also reported that the alarm messages could not be controlled despite the hospital staff making adjustments to the driver settings.The customer also reported that the patient was subsequently switched to the backup companion 2 driver.There was no reported adverse patient impact.The companion 2 driver was returned to syncardia for evaluation.Visual inspection of the driver's external components revealed no anomalies.The patient file was copied and reviewed.Multiple left pressure incorrect alarms were observed.During these alarms, the left pressure was consistently higher than the set point, which confirmed the customer-reported issue.During investigation testing, the customer-reported left pressure incorrect alarm was duplicated.The root cause was a malfunction of the left electronic pressure regulator.The left electronic pressure regulator was taken out of service and replaced.The issues with electronic pressure regulators in relation to pressure incorrect alarms are being investigated in a capa (corrective or preventive action).The capa will document the investigation including, but not limited to, potential root cause and corrective actions associated with compressor cycling because of a malfunction of the manual pressure regulator.Despite the customer-reported left pressure incorrect alarms, risk to the patient was low because the driver continued to perform its life-sustaining functions and the cardiac output remained within the acceptable 5 to 8 l/min (liters per minute).The companion 2 driver was serviced and passed all functional and performance testing before being placed into finished goods.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.
|