STRYKER ORTHOPAEDICS-MAHWAH UNIVERSAL DRIVER; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
Catalog Number 6541-4-801 |
Device Problems
Disconnection (1171); Insufficient Information (3190)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/01/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
When completed, the investigation results will be submitted in a supplemental report.
|
|
Event Description
|
Nurse (b)(6) reported that the triathlon adapter did not hold the instruments, they came off.
|
|
Manufacturer Narrative
|
An event regarding an adapter not holding mating parts involving a universal driver was reported.The event was not confirmed.Method & results: device evaluation and results: visual and functional testing were performed and although the device does display minor wear it was found to be fully functional.Medical records received and evaluation: not performed as patient factors did not contribute to the event.Device history review: review of the device history records indicates that all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for this lot.Conclusions: the investigation could not determine the root cause of the event as functional testing of the device found it to be fully functional.There is no indication at this time that the design, materials, or manufacturing of the subject device contributed to the event.If additional information becomes available, this investigation will be reopened.
|
|
Event Description
|
Nurse (b)(6) reported that the triathlon adapter did not hold the instruments, they came off.
|
|