Catalog Number 6541-4-801
Device Problems Disconnection (1171); Insufficient Information (3190)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 10/01/2015
Event Type  malfunction  
Manufacturer Narrative
When completed, the investigation results will be submitted in a supplemental report.
 
Event Description
Nurse (b)(6) reported that the triathlon adapter did not hold the instruments, they came off.
 
Manufacturer Narrative
An event regarding an adapter not holding mating parts involving a universal driver was reported.The event was not confirmed.Method & results: device evaluation and results: visual and functional testing were performed and although the device does display minor wear it was found to be fully functional.Medical records received and evaluation: not performed as patient factors did not contribute to the event.Device history review: review of the device history records indicates that all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for this lot.Conclusions: the investigation could not determine the root cause of the event as functional testing of the device found it to be fully functional.There is no indication at this time that the design, materials, or manufacturing of the subject device contributed to the event.If additional information becomes available, this investigation will be reopened.
 
Event Description
Nurse (b)(6) reported that the triathlon adapter did not hold the instruments, they came off.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL DRIVER
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla navedo
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5185017
MDR Text Key30220184
Report Number0002249697-2015-03510
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
PMA/PMN Number
K123486
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 10/01/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6541-4-801
Device Lot NumberN3M30
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/23/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/11/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/08/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-