Manufacturing site evaluation: samples received: there are no samples available.Analysis and results: there are no previous complaints of this code batch.There are no units in stock.(b)(4) units were manufactured and distributed.Without any closed sample an analysis cannot be carried out in order to make a decision.As indicated in the dafilon instructions for use: "when working with suture materials great care should be taken to ensure that the use of surgical instruments, such as tweezers and needle holders, does not cause the material to be damaged by being pinched or kinked".Final conclusion: without samples, not in position of studying if the affected product does not fulfill the oem specifications.Complaint is not justified.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to take an actions on the product.Corrective/preventive actions: according to the internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
|
Country of complaint: (b)(4).The thread breaks extremely easily, forming at the break a fraying of the thread and the thread slides on the first knot, making it difficult for the second knot.Also, the wire is very elastic and the shape memory is excessively marked.The thread is damaged.
|