Catalog Number TC003 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/05/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Conclusion: the product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.In addition, a review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
|
|
Event Description
|
It was reported that the patient underwent an endometrial thermal ablation procedure on (b)(6) 2015.During the procedure, the balloon had detached from the catheter after removing the fluid and removing the wand from the patient.The balloon was retrieved from the patient using stone forceps without additional dissection.There was no consequence to the patient.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation and the balloon was burst.The balloon was received inside a plastic bag.It is possible that the fluid inside the catheter was not completely removed, causing the balloon to burst.This was caused as a result of severe non-normal handling by the user.
|
|