Catalog Number 2C8750
Device Problem Occlusion Within Device (1423)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a clearlink blood set experienced clotting at the end of the filter and the blood would not drip at the proper rate.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).It was reported that the facility had not been inverting the filter in order to prime the set correctly which had been causing the blood clots.The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCESS
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE - HAINA, SAN CRISTOBAL
parque industrial itabo, zona
haina
DR 
Manufacturer (Section G)
BAXTER HEALTHCARE - HAINA, SAN CRISTOBAL
parque industrial itabo, zona
haina
DR  
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5192922
MDR Text Key30015303
Report Number1416980-2015-40586
Device Sequence Number1
Product Code BRZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K993120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number2C8750
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/03/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown