Catalog Number 511.730
Device Problem Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that the chuck with key device could not be attached to the device anymore.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was confirmed.It was determined that the device was wornout and would no longer go into the machine.The assignable root cause was due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHUCK W/KEY F/COMPACT AIR DRIVE+POWER DR
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ  CH4436
Manufacturer (Section G)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ   CH4436
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5197368
MDR Text Key30595629
Report Number8030965-2015-11759
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number511.730
Device Lot Number1178074
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received12/15/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/23/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-