• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS BURR-ATTACHM MEDIUM F/EPD+APD; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESSORIES AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS BURR-ATTACHM MEDIUM F/EPD+APD; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESSORIES AND ATTACHMENT Back to Search Results
Catalog Number 05.001.046
Device Problems Unstable (1667); Mechanical Jam (2983); Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the bearing seized,jammed and heavy moving.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to normal wear.If information is obtained that was not available for the initial medwatch, a supplemental medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that the attachment device could not be positioned when mounted to a device.According to the reporter the part was out of balance and could not be used.Durin in-house engineering evaluation, it was observed that the bearings seized, jammed and were heavy moving.It was not reported if the device was used in surgery, or if there was patient involvement.There were no delays in a scheduled surgical procedure.A spare device was not available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BURR-ATTACHM MEDIUM F/EPD+APD
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESSORIES AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5198713
MDR Text Key30569861
Report Number3009450871-2015-14014
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
PK043310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/07/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.046
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/23/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/16/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-