DEPUY SYNTHES POWER TOOLS BURR-ATTACHM MEDIUM F/EPD+APD; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESSORIES AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.046 |
Device Problems
Unstable (1667); Mechanical Jam (2983); Positioning Problem (3009)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the bearing seized,jammed and heavy moving.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to normal wear.If information is obtained that was not available for the initial medwatch, a supplemental medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported from (b)(6) that the attachment device could not be positioned when mounted to a device.According to the reporter the part was out of balance and could not be used.Durin in-house engineering evaluation, it was observed that the bearings seized, jammed and were heavy moving.It was not reported if the device was used in surgery, or if there was patient involvement.There were no delays in a scheduled surgical procedure.A spare device was not available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|