• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - IRVINE INFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - IRVINE INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number J2C1009
Device Problem Cut In Material (2454)
Patient Problem No Patient Involvement (2645)
Event Date 10/19/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that there was a small cut on the tube of a large volume infusor.The reporter stated that the cut was 3cm from the connection to the stress member.The device was filled with saline.This was noted before patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
(b)(4).The device was manufactured january 7, 2015 ¿ january 8, 2015.The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional flow test was performed, and no signs of a leak were observed from the tubing.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5199190
MDR Text Key30324446
Report Number1416980-2015-40860
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue NumberJ2C1009
Device Lot Number15A003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received11/04/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-