(b)(4).A review of the information in the complaint file indicates this investigation was performed by a covidien/medtronic technical center for the reported condition of; damaged power cord.Therefore, this report will be based on information provided by the service center.The unit was triaged and the complaint was confirmed; power cord is damaged with exposed copper wire.The root cause of the power cord failure can be attributed to rough handling.Power cords periodically require replacement due to age, usage and user damage.The power cord was replaced to correct the problem.The unit was then fully tested.Scd 700 compression system was manufactured in 2013.A review of the device history record shows this device was released meeting all manufacturing specifications.
|