• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN, FORMERLY USSC PUERTO RICO INC CLEARIFY VISUALIZATION SYSTEM; ANTI FOG SOLUTION AND ACCESSORIES, ENDOSCOPY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN, FORMERLY USSC PUERTO RICO INC CLEARIFY VISUALIZATION SYSTEM; ANTI FOG SOLUTION AND ACCESSORIES, ENDOSCOPY Back to Search Results
Model Number 21345
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
According to the reporter, the covering of the hub's bottom is fraying - it is not a clear cut of the edges.The consequence is that some small pieces of the "paper" has been laying on the or table.This could create the situation where these "paper pieces" could potentially fall inside the patient.No patient injured, no extension of incision or surgery time, no blood loss, no tissue damage, no change of procedure, nothing fell into cavity, no reinforcement material used.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of two photographs of a device opened by the account.This evaluation was based on a technical review of all data received from the site, a pmv review of manufacturing records, a pmv review of complaint trends, engineering review of the product and an evaluation of the returned device.The visual inspection of the photograph demonstrates a frayed covering on the bottom of the device.Based on the evaluation of the returned photos engineering was unable to confirm the reported condition.A review of the device history record could not be performed because the lot number was not provided.However, records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality specifications at the time of manufacture.Subsequently, the complaint data did not display an increased trend.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARIFY VISUALIZATION SYSTEM
Type of Device
ANTI FOG SOLUTION AND ACCESSORIES, ENDOSCOPY
Manufacturer (Section D)
COVIDIEN, FORMERLY USSC PUERTO RICO INC
building 911-67
sabanetas industrial park
ponce PR 00731
Manufacturer (Section G)
COVIDIEN, FORMERLY USSC PUERTO RICO INC
building 911-67
sabanetas industrial park
ponce PR 00731
Manufacturer Contact
sharon murphy
60 middletown ave
north haven, CT 06473
2034925267
MDR Report Key5200753
MDR Text Key30619392
Report Number2647580-2015-00774
Device Sequence Number1
Product Code OCT
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
K062779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 10/28/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21345
Device Catalogue Number21345
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/22/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-