COVIDIEN, FORMERLY USSC PUERTO RICO INC CLEARIFY VISUALIZATION SYSTEM; ANTI FOG SOLUTION AND ACCESSORIES, ENDOSCOPY
|
Back to Search Results |
|
Model Number 21345 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
According to the reporter, the covering of the hub's bottom is fraying - it is not a clear cut of the edges.The consequence is that some small pieces of the "paper" has been laying on the or table.This could create the situation where these "paper pieces" could potentially fall inside the patient.No patient injured, no extension of incision or surgery time, no blood loss, no tissue damage, no change of procedure, nothing fell into cavity, no reinforcement material used.
|
|
Manufacturer Narrative
|
Evaluation summary: post market vigilance (pmv) led an evaluation of two photographs of a device opened by the account.This evaluation was based on a technical review of all data received from the site, a pmv review of manufacturing records, a pmv review of complaint trends, engineering review of the product and an evaluation of the returned device.The visual inspection of the photograph demonstrates a frayed covering on the bottom of the device.Based on the evaluation of the returned photos engineering was unable to confirm the reported condition.A review of the device history record could not be performed because the lot number was not provided.However, records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality specifications at the time of manufacture.Subsequently, the complaint data did not display an increased trend.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.
|
|
Search Alerts/Recalls
|
|
|