• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE VERSACARE BED; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE VERSACARE BED; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number 3200
Device Problems Device Alarm System (1012); Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/16/2015
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the external alarm assembly inoperable per the hill-rom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Brake casters should be checked for cuts, wear and quality of tread, etc.And replaced when necessary.Examine the brakes to see whether the bed moves when the brake are set.Replace as necessary.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the external alarm to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the bed exit alarm was inoperative.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
john cummings
1069 state route 46 east
batesville, IN 47006
8129312869
MDR Report Key5207091
MDR Text Key30779334
Report Number1824206-2015-01013
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received10/16/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-