CODMAN AND SHURTLEFF, INC ORBIT GALAXY DETACHABLE COIL SYSTEM; ARTIFICIAL EMBOLIZATION DEVICE
|
Back to Search Results |
|
Catalog Number 640CR1230 |
Device Problem
Delivery System Failure (2905)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/14/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: prowler lpes microcatheter (606-s155fx/17259727).(b)(4).The device was returned for analysis; however, the analysis has not yet been completed.Additional information will be submitted within 30 days of receipt.This is 1 of 2 mdr reports being submitted for this complaint with associated report numbers of 3008264254-2015-00071 and 3008264254-2015-00070.
|
|
Event Description
|
As reported via a healthcare professional, during coil embolization of a right internal carotid artery cavernous segment aneurysm, an orbit galaxy coil (640cr1230/16054124) pushed the prowler lpes microcatheter (606-s155fx/17259727) out of the aneurysm due to coil positioning difficulty and difficulty in getting the coil to conform to the aneurysm wall, and the coil was not able to be deployed.There was no coil protrusion from the aneurysm or reduced blood flow due to the coil positioning difficulty, and no resistance between the coil and microcatheter or need to withdraw the microcatheter with the coil.Another orbit galaxy coil (640cr1230/16054124) hypotube ¿bent¿ while being loaded into the microcatheter, and when the doctor tried to straighten out the kink by applying force, it ¿snapped¿.The coil had not exited the microcatheter and was removed from the microcatheter.The same microcatheter was used to implant all 48 coils implanted during the procedure.There had been no resistance between the coils and microcatheter and no damage to the actual coil.It was reported that the devices had been used and prepped as per the instructions for use (ifu), and had appeared normal prior to use.The events did not result in patient injury or clinically significant delay in the procedure.
|
|
Manufacturer Narrative
|
Complaint conclusion: as reported via a healthcare professional, during coil embolization of a right internal carotid artery cavernous segment aneurysm, an orbit galaxy coil (640cr1230/16054124) pushed the prowler lpes microcatheter (606-s155fx/17259727) out of the aneurysm due to coil positioning difficulty and difficulty in getting the coil to conform to the aneurysm wall, and the coil was not able to be deployed.There was no coil protrusion from the aneurysm or reduced blood flow due to the coil positioning difficulty, and no resistance between the coil and microcatheter or need to withdraw the microcatheter with the coil.Another orbit galaxy coil (640cr1230/16054124) hypotube ¿bent¿ while being loaded into the microcatheter, and when the doctor tried to straighten out the kink by applying force, it ¿snapped¿.The coil had not exited the microcatheter and was removed from the microcatheter.The same microcatheter was used to implant all 48 coils implanted during the procedure.There had been no resistance between the coils and microcatheter and no damage to the actual coil.It was reported that the devices had been used and prepped as per the instructions for use (ifu), and had appeared normal prior to use.The events did not result in patient injury or clinically significant delay in the procedure.A non-sterile orbit galaxy cmplx frame coil 12x30 was received coiled inside of a plastic bag.The delivery tube was found broken as well as a kinked condition was noted on it.The introducer was found zipped without damage.The support coil, gripper and the embolic coil were found inside of the introducer.The gripper, embolic coil and the hypotube were inspected under microscope; no damages were noted on the gripper and the embolic coil while the edge of the broken section of the hypotube show evidence that appear that it was kinked before it was broken.The review of lot 16054124 revealed 1 defect for kink /bent and it could be related to the reported complaint; however; the dhr review confirmed that the rejected units were properly segregated and discarded.Controls are in place to detect such nonconformities.No other issues were noted that were considered potentially related to the reported complaint.The device was returned for analysis.The delivery tube bent/fractured was confirmed.The cause of the kinked and the separated condition was apparently caused by applying excessive force on it but it could not be conclusively determined.Since there was no evidence of a manufacturing issue related to the event, no corrective actions will be taken at this time.This is 1 of 2 mdr reports being submitted for this complaint with associated report numbers of 3008264254-2015-00071 and 3008264254-2015-00070.
|
|
Search Alerts/Recalls
|
|
|