• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLAINFIELD SCD EXPRESS COMP SYSTEM JAPAN; SCD CONTROLLER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLAINFIELD SCD EXPRESS COMP SYSTEM JAPAN; SCD CONTROLLER Back to Search Results
Model Number 95257
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 11/04/2015
Event Type  malfunction  
Manufacturer Narrative
Submit date: 11/06/2015.An investigation is currently underway.Upon completion, a full detailed investigation will be provided.
 
Event Description
On (b)(6) 2015 the customer initiated a service repair request regarding an issue with the power cord.The service tech found the unit had a damaged power cord with exposed copper wires.No patient involvement.
 
Manufacturer Narrative
A review of the information in the complaint file indicates this investigation was performed by a medtronic technical center for the reported condition of; damaged power cord.Therefore, this report will be based on information provided by the technical center.The unit was triaged and the complaint was confirmed.The root cause of the power cord failure can be attributed to rough handling of the unit.Power cords periodically require replacement due to age, usage and user damage.The service manual instructs the user to periodically inspect the power cords resistance to ensure its electrical safety.The power cord was replaced to correct the problem.Scd express compression system was manufactured in 2007.A review of the device history record shows this device was released meeting all manufacturing specifications.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.Based on the investigation, no corrective action is needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCD EXPRESS COMP SYSTEM JAPAN
Type of Device
SCD CONTROLLER
Manufacturer (Section D)
PLAINFIELD
2824 airwest boulevard
plainfield IN 46168
Manufacturer (Section G)
COVIDIEN
2824 airwest blvd
plainfield IN 46168
Manufacturer Contact
thom mcnamara
15 hampshire street
mansfield, MA 02048
5084524811
MDR Report Key5207826
MDR Text Key30880469
Report Number3008361498-2015-00014
Device Sequence Number1
Product Code JOW
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number95257
Device Catalogue Number95257
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received11/04/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-