DEPUY SYNTHES POWER TOOLS 5 CM HEAVY DUTY SHORT ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT
|
Back to Search Results |
|
Catalog Number SHORT-HD-G1 |
Device Problem
Component Falling (1105)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/22/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional narrative: the actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the bearing stack had come out of the motor.Therefore, the reported condition was confirmed.The assignable root cause was determined to be insufficient loctite adhesive applied to the thread of the bearing stack during the manufacturing process, improper assembly/manufacture.This issue has been escalated to capa.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that during pre-surgery, it was observed that the bearings fell out of the attachment device.There was no delay due to the planned surgical procedure as an identical spare device was available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|