(b)(4).Device evaluated by manufacturer - the returned product consisted of and angiojet solent omni thrombectomy system.Visual and tactile examinations showed that there was a break in the hypo tube approximately 56cm from the tip of the catheter.The thrombectomy system was inserted in to the ultra console for functional testing.The solent omni would not prime.The batch number is unknown and the manufacturing records for the complaint device could not be reviewed.The most probable root cause is considered handling damage as the event occurred prior to patient contact.(b)(4).
|