MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/21/2015
Event Type  malfunction  
Manufacturer Narrative
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
 
Event Description
It was reported that during a cryo ablation procedure when the sheath was in the left atrium of the patient, the stopcock detached from the sheath extension while flushing and aspirating the sheath.The sheath was replaced and the procedure was completed with cryo.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: data files and a photo were returned and analyzed.The data files did not show any system notices on the re ported event date.The pictures provided by the facility demonstrate the stopcock distal end luer was completely detached from the side port tube proximal end.No product was returned for investigation.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
16771 chemin ste-marie
kirkland H9H 5 H3
CA  H9H 5H3
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5213726
MDR Text Key31120021
Report Number3002648230-2015-00396
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/13/2017
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number66037
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/19/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/14/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-