Model Number 4FC12 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/21/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
|
|
Event Description
|
It was reported that during a cryo ablation procedure when the sheath was in the left atrium of the patient, the stopcock detached from the sheath extension while flushing and aspirating the sheath.The sheath was replaced and the procedure was completed with cryo.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: data files and a photo were returned and analyzed.The data files did not show any system notices on the re ported event date.The pictures provided by the facility demonstrate the stopcock distal end luer was completely detached from the side port tube proximal end.No product was returned for investigation.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|