• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MOUNTAIN HOME MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MOUNTAIN HOME MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number R5C4482
Device Problems Leak/Splash (1354); Loose or Intermittent Connection (1371); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/02/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a minicap transfer set leaked during patient use.The reporter stated that the patient had closed the roller sleeve and heard a noise, then disconnected the solution bag.The transfer set had then leaked.While the transfer set was being replaced, the roller sleeve came loose from the set.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).The device was received for evaluation.A visual inspection with magnification noted broken occluder feet.Underwater leak testing, clear passage testing, and clamp function testing were performed.Underwater leak testing and clamp function testing noted a leak through the set.The reported condition was verified.The cause of the leak was determined to be broken occluder feet.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP TRANSFER SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5214034
MDR Text Key31087508
Report Number1416980-2015-41501
Device Sequence Number1
Product Code KDJ
Combination Product (y/n)N
Reporter Country CodeNL
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 11/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberR5C4482
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/21/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-