BAXTER HEALTHCARE - MOUNTAIN HOME MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number R5C4482 |
Device Problems
Leak/Splash (1354); Loose or Intermittent Connection (1371); Noise, Audible (3273)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/02/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a minicap transfer set leaked during patient use.The reporter stated that the patient had closed the roller sleeve and heard a noise, then disconnected the solution bag.The transfer set had then leaked.While the transfer set was being replaced, the roller sleeve came loose from the set.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
(b)(4).The device was received for evaluation.A visual inspection with magnification noted broken occluder feet.Underwater leak testing, clear passage testing, and clamp function testing were performed.Underwater leak testing and clamp function testing noted a leak through the set.The reported condition was verified.The cause of the leak was determined to be broken occluder feet.Should additional relevant information become available, a supplemental report will be submitted.
|
|