Catalog Number AI-07124 |
Device Problem
Defective Component (2292)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/08/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that while in the cath lab the wedge pressure catheter was inserted via the patient's femoral artery.During use the user found the "hub end" was defective.As a result, the wedge pressure catheter was removed and replaced.There was no reported patient death, injury or complications.Medical / surgical intervention was not required.There was a delay / interruption for the time frame required to replace the (b)(4).
|
|
Manufacturer Narrative
|
(b)(4).Device evaluation: a 5fr wedge pressure catheter was returned for evaluation.No damage, abnormalities, or kinks were noted to the catheter.The capacity of the balloon was noted to be 0.75cc.The syringe typically supplied with the kit was not returned with the catheter.The balloon was visually inspected and appeared to be typical upon initial examination.The hubs of the injection and inflation lumen appeared intact.No kinks were noted to the catheter body.The inflation lumen was injected with 0.75cc of air using a lab inventory syringe.The balloon was fully inflated.The balloon deflated in less than 3 seconds when the syringe was removed per specification.The balloon held inflation for at least one minute.The balloon inflated symmetrically.The injection lumen was aspirated and flushed with air.There was slight resistance during aspiration.A small spec of blood exited from the catheter's injection lumen after being flushed with water.After clearing the clot, the catheter was able to be aspirated and flushed with no resistance.A device history record review was not performed.There was no confirmed product failure with the returned sample.Other remarks: conclusion: the reported complaint that the hub was defective is not confirmed.No visual or functional issues were noted to any of the hubs on the device.The device passed functional testing.The root cause of the complaint is undetermined.
|
|
Event Description
|
It was reported that while in the cath lab the wedge pressure catheter was inserted via the patient's femoral artery.During use the user found the "hub end" was defective.As a result, the wedge pressure catheter was removed and replaced.There was no reported patient death, injury or complications.Medical / surgical intervention was not required.There was a delay / interruption for the time frame required to replace the (b)(4).
|
|
Search Alerts/Recalls
|