Model Number M51PSEMIBLUE |
Device Problems
Migration or Expulsion of Device (1395); Device Displays Incorrect Message (2591)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional information become available a supplemental record will be filed.
|
|
Event Description
|
Per dealer the error code e503 came on joystick, right brake fault,.
|
|
Manufacturer Narrative
|
Product was returned for evaluation.The return fields in (b)(4) state: the underlying cause documented on the irs or return fields in (b)(4) is identified as: motor 1162493 / brake failure-electrical / electrical brake failure - bench test.Wiggle wires by the strain relief and it will shut off and give a brake fault during bench test.Product received expanded evaluation.The expanded evaluation report states: utilizing existing complaint information, actual observations, and functional testing of the returned product in its "as received" condition, the complaint was confirmed for the left motor and gearbox of having a five flash error code indicating a left park brake fault caused by strain relief pullout.Complaint was confirmed.The underlying cause is strain relief pullout.Root cause could not be determined after reviewing the documentation in this investigation.
|
|
Event Description
|
Product was returned for evaluation.The return fields in oracle state: the underlying cause documented on the (b)(4) or return fields in (b)(4) is identified as: motor 1162493 / brake failure-electrical / electrical brake failure - bench test.Wiggle wires by the strain relief and it will shut off and give a brake fault during bench test.Product received expanded evaluation.The expanded evaluation report states: utilizing existing complaint information, actual observations, and functional testing of the returned product in its "as received" condition, the complaint was confirmed for the left motor and gearbox of having a five flash error code indicating a left park brake fault caused by strain relief pullout.
|
|
Search Alerts/Recalls
|