MAUDE Adverse Event Report: INVACARE REHABILITATION EQUIPMENT CO. POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI; WHEELCHAIR, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVACARE REHABILITATION EQUIPMENT CO. POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI; WHEELCHAIR, POWERED Back to Search Results
Model Number M51PSEMIBLUE
Device Problems Migration or Expulsion of Device (1395); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Should additional information become available a supplemental record will be filed.
 
Event Description
Per dealer the error code e503 came on joystick, right brake fault,.
 
Manufacturer Narrative
Product was returned for evaluation.The return fields in (b)(4) state: the underlying cause documented on the irs or return fields in (b)(4) is identified as: motor 1162493 / brake failure-electrical / electrical brake failure - bench test.Wiggle wires by the strain relief and it will shut off and give a brake fault during bench test.Product received expanded evaluation.The expanded evaluation report states: utilizing existing complaint information, actual observations, and functional testing of the returned product in its "as received" condition, the complaint was confirmed for the left motor and gearbox of having a five flash error code indicating a left park brake fault caused by strain relief pullout.Complaint was confirmed.The underlying cause is strain relief pullout.Root cause could not be determined after reviewing the documentation in this investigation.
 
Event Description
Product was returned for evaluation.The return fields in oracle state: the underlying cause documented on the (b)(4) or return fields in (b)(4) is identified as: motor 1162493 / brake failure-electrical / electrical brake failure - bench test.Wiggle wires by the strain relief and it will shut off and give a brake fault during bench test.Product received expanded evaluation.The expanded evaluation report states: utilizing existing complaint information, actual observations, and functional testing of the returned product in its "as received" condition, the complaint was confirmed for the left motor and gearbox of having a five flash error code indicating a left park brake fault caused by strain relief pullout.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI
Type of Device
WHEELCHAIR, POWERED
Manufacturer (Section D)
INVACARE REHABILITATION EQUIPMENT CO.
no.5 weixi road, sip
suzhou jiangsu 21512 1
CH  215121
Manufacturer (Section G)
INVACARE REHABILITATION EQUIPMENT CO.
no.5 weixi road, sip
suzhou jiangsu 21512 1
CH   215121
Manufacturer Contact
kevin guyton
one invacare way
elyria, OH 44035
8003336900
MDR Report Key5217587
MDR Text Key31141961
Report Number3008262382-2015-01928
Device Sequence Number1
Product Code ITI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 11/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberM51PSEMIBLUE
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received11/23/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-