During high frequency oscillatory ventilation, the respiratory care practitioner (rcp) noted that the indicator for deltap climbed from 30 delta p to 140 deltap on the front panel indicator and the ventilator began alarming.The rcp could not control the climbing deltap indication by correcting power or pressure adjustments.The patient airway pressure remained at 20.5cm h20 and the patient's chest movements seemed "normal".Biomedical was contacted to assist in exchanging the ventilator and the problem ventilator was sent to biomedical for repair failure analysis.Bmet troubleshooting noted fluid buildup and pooling near the internal differential pressure transducer and static pneumatic tubing loop.Bmet noted that this was the third failure of this type in the last 4 months.Failure was noted on the adult version of the oscillator ventilator.Bmet cleared and flushed the tubing and transducer and retested the oscillator to normal operation and performance.Bmet contacting manufacturer with failure details for additional investigation.Manufacturer response for oscillatory ventilator, sensormedics (per site reporter): awaiting manufacturer response.
|