The customer questioned low results for 20 patients tested for cancer antigen 125 (ca 125 ii) after changing a reagent pack.Results were provided for 2 patients.Of the data provided, the ca 125 ii results for 1 patient were erroneous and reported outside of the laboratory.The initial ca 125 ii result was 12.0 u/ml.After performing a reagent pack calibration, the repeat result was 189 u/ml.The repeat result was considered to be correct.Both results were reported to the patient's physician.No adverse event occurred.The ca 125 ii reagent lot number was 181804.The expiration date was not provided.It was noted that the customer performed a lot calibration on (b)(6) 2015 and the signals were high.After the calibration on (b)(6) 2015, the customer performed reagent pack calibration several times but did not perform a lot calibration.A reagent pack calibration was performed on (b)(6) 2015 and the values were higher than expected but acceptable.A specific root cause could not be identified.Additional information for further investigation was requested but was not provided.A general reagent and instrument issue can be excluded.The most likely root cause of the erroneous results was the calibration that was performed on (b)(6) 2015.After the calibration that was performed on (b)(6) 2015, the customer did not observe any more questionable low results.
|