A healthcare facility in (b)(6), reported an incident to a fisher & paykel healthcare (fph) representative where a patient was receiving heliox and supplemental oxygen in a set up which included an optiflow junior interface, rt330 optiflow breathing circuit and an mr850 respiratory humidifier.It was further reported that at some point during the night the staff turned off the gas flow to the mr850 respiratory humidifier.They further reported that the humidification chamber on the mr850 deformed.The customer further informed us that the reported incident was due to an "operator error".No patient consequence was reported.
|