MAUDE Adverse Event Report: BAXTER HEALTHCARE - MOUNTAIN HOME MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MOUNTAIN HOME MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number R5C4482
Device Problem Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).It was reported that this happened approximately eight weeks before (b)(6) 2015.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a minicap transfer set had a defective rotatory clamp which had been over-rotated and did not close correctly.The patient may have applied excessive force and twisted the clamp in the wrong direction.The patient had used an unspecified device to open the set and had cleaned the transfer set with octenisept.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP TRANSFER SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5228120
MDR Text Key31324726
Report Number1416980-2015-40128
Device Sequence Number1
Product Code KDJ
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K882498
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberR5C4482
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/17/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-