MAUDE Adverse Event Report: CODMAN & SHURTLEFF, INC. CODMAN DISPOSABLE PERFORATOR; DRILLS, BURRS, TREPHINES & ACC.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CODMAN & SHURTLEFF, INC. CODMAN DISPOSABLE PERFORATOR; DRILLS, BURRS, TREPHINES & ACC. Back to Search Results
Catalog Number 26-1221
Device Problems Device Issue (2379); Dull, Blunt (2407)
Patient Problem No Information (3190)
Event Date 09/09/2014
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Upon completion of the investigation, a follow up report will be filed.
 
Event Description
The perforator was not functioning because it was not cutting enough.There was no clinical consequence.Another perforator was used.
 
Manufacturer Narrative
Upon completion of the investigation it was noted that the customer¿s complaint of "perforator was not cutting enough" was not verified.This perforator met the test method acceptance requirements; proper engagement and disengagement were achieved with every drill hole.There was no erratic or poor cutting action.The device history records (perforator assembly) show all tests and inspections, including a drilling test on each perforator, met specification requirements.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOSABLE PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACC.
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC.
new bedford industrial park
new bedford MA 02745
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key5228178
MDR Text Key31336368
Report Number1226348-2015-10677
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2015
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number26-1221
Device Lot NumberGH015S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received12/10/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/23/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-