• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAUSCH + LOMB CRYSTALENS ACCOMMODATING IOL; LENS, INTRAOCULAR, ACCOMMODATIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB CRYSTALENS ACCOMMODATING IOL; LENS, INTRAOCULAR, ACCOMMODATIVE Back to Search Results
Model Number AT50AO
Device Problem Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/20/2015
Event Type  Injury  
Manufacturer Narrative
Investigation is underway.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
It was reported that the lens was implanted and removed intraoperatively due to the surgeon's dissatisfaction with the lens positioning in the eye.The incision was enlarged and the lens cut out of the eye.The backup lens was implanted and sutures were used.Additional information has been requested but has not been received.
 
Manufacturer Narrative
The lens was returned for evaluation in three pieces.Visual inspection of the lens found tears and a portion of the leading haptic and optic missing.Further testing cannot be performed due to the damage.Inspection on a retain sample from the same lot was performed with all dimensional measurements found to be within specification.The lot history record was reviewed and there were no non-conformities or anomalies related to this complaint.Based on the information provided, we are unable to determine a root cause.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRYSTALENS ACCOMMODATING IOL
Type of Device
LENS, INTRAOCULAR, ACCOMMODATIVE
Manufacturer (Section D)
BAUSCH + LOMB
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB
10574 acacia street suite d1
rancho cucamonga CA 91730
Manufacturer Contact
sharon spencer
50 technology drive west
irvine, CA 92618
9493985698
MDR Report Key5229826
MDR Text Key31386578
Report Number0001313525-2015-02955
Device Sequence Number1
Product Code NAA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2019
Device Model NumberAT50AO
Device Lot Number7463402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/16/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-